A minimum source-skin-distance of less than 30 cm and not identifying the tube manufacturer on the tube housing label resulted in the SP-HF-4.0 Portable Systems not complying with the U.S. Federal performance standard.
Source facts
Action
An 'Important Safety Notice' dated 10/27/03 was sent to the one direct account on the same date, informing them of the non-compliances with the performance standard and requesting them to call Sedecal at 800-920-9525 to arrange for the modification of the x-ray unit at no cost to the user.
Initiated
2003-10-27
Root cause
Other
Terminated
2004-03-24
Distribution
Vermont
Recall number
Z-0001-04
Recall status
Terminated
CDRH record ID
30043
FDA classified
2003-10-29
Recall event number
27549