Failure-- product can kink or ovalize, which may result in difficulty in catheter advancement or delivery of other devices and damage to catheter, which could lead to patient injury or death as a result of portions being retained in the patient.
Source facts
Action
The firm issued an Urgent Voluntary Field Removal Notice letter on 09/22/2009 to its consignees, asking them to inspect their stocks; remove affected product and place in quarantine for return to Penumbra Inc.; and complete and return the attached product identification / return form.
Initiated
2009-09-22
Root cause
Device Design
Terminated
2010-11-24
Distribution
Worldwide Distribution -- USA, Canada, Hong Kong, Chile, the UK, Denmark, Norway, Sweden, Finland, Netherlands, Germany, Switzerland, Austria, and Spain.
Recall number
Z-0001-2010
Recall status
Terminated
CDRH record ID
85486
FDA classified
2009-10-05
Recall event number
53426