Open original FDA record ↗Recall record
- FDA identifier
- Z-0001-2015
- Company in FDA record
- Invivo Corporation
- FDA classification date
- Oct 1, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
During production, excess metal (aluminum) debris was found stuck to the cling plastic that covers the device's pole assembly. The metallic debris could affect system performance due to the potential for a failure to power up or loss of power. This could result in the loss of monitoring and a delay in diagnosis of the patient.
Source facts
- Action
- The firm, Philips Healthcare, sent an "URGENT- Medical Device Recall" letter dated June 17, 2014, to its customers informing them of the product, problem and actions taken. The firm informed the customers that there are no additional actions required on their part. The Philips/Invivo Account Manager will contact them in order to arrange for replacement of the device. If you need any further information or support concerning this issue, contact your local Philips/Invivo representative: 877-468-4861 option 1, then option 2.
- Initiated
- 2014-09-08
- Root cause
- Material/Component Contamination
- Terminated
- 2017-01-09
- Distribution
- Worldwide Distribution: US distribution in state of: CA and countries of: Poland, Singapore, and Sweden.
- Recall number
- Z-0001-2015
- Recall status
- Terminated
- CDRH record ID
- 129828
- FDA classified
- 2014-10-01
- Recall event number
- 69192