Open original FDA record ↗Recall record
- FDA identifier
- Z-0002-2008
- Company in FDA record
- Beckman Coulter Inc
- FDA classification date
- Oct 16, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Pressure (from recapped tubes) or vacuum (from under-filled short-draw tubes) will cause inaccurate results if not properly vented.
Source facts
- Action
- A Product Corrective Action (PCA) letter was mailed on Feb 22, 2007, to all FP1000 customers to inform them of an issue associated with the FP1000 Cell Preparation System. Pressure (from recapped tubes) or vacuum (from under-filled short-draw tubes) will cause inaccurate results if not properly vented. Under these conditions the FP1000 will under-dispense specimen volume leading to lower than expected absolute counts. Specimen tubes that are recapped or underfilled must therefore be vented prior to running on the FP1000. If the initial state of the specimen tube is unknown, then it must be vented to ensure accurate results prior to running on the FP1000. Additional instructions are provided. A Fax Back Response Form was also provided.
- Initiated
- 2007-02-22
- Root cause
- Labeling design
- Terminated
- 2012-04-26
- Distribution
- Nationwide (MA, IN, IL, UT, RI, NJ), and Canada.
- Recall number
- Z-0002-2008
- Recall status
- Terminated
- CDRH record ID
- 53495
- FDA classified
- 2007-10-16
- Recall event number
- 38369