Z-0002-2008: FP1000 Cell Preparation System Software versions 1.1 & 1.2, Part Number 624922, for in vitro diagnostic u

Recall record Z-0002-2008 for Beckman Coulter Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0002-2008
Company in FDA record
Beckman Coulter Inc
FDA classification date
Oct 16, 2007
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Pressure (from recapped tubes) or vacuum (from under-filled short-draw tubes) will cause inaccurate results if not properly vented.

Source facts

Action
A Product Corrective Action (PCA) letter was mailed on Feb 22, 2007, to all FP1000 customers to inform them of an issue associated with the FP1000 Cell Preparation System. Pressure (from recapped tubes) or vacuum (from under-filled short-draw tubes) will cause inaccurate results if not properly vented. Under these conditions the FP1000 will under-dispense specimen volume leading to lower than expected absolute counts. Specimen tubes that are recapped or underfilled must therefore be vented prior to running on the FP1000. If the initial state of the specimen tube is unknown, then it must be vented to ensure accurate results prior to running on the FP1000. Additional instructions are provided. A Fax Back Response Form was also provided.
Initiated
2007-02-22
Root cause
Labeling design
Terminated
2012-04-26
Distribution
Nationwide (MA, IN, IL, UT, RI, NJ), and Canada.
Recall number
Z-0002-2008
Recall status
Terminated
CDRH record ID
53495
FDA classified
2007-10-16
Recall event number
38369

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗