Open original FDA record ↗Recall record
- FDA identifier
- Z-0002-2009
- Company in FDA record
- GE Medical Systems, LLC
- FDA classification date
- Sep 24, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
GE Healthcare has recently become aware of a potential risk of observing a blank image after exposure on Revolution XR/d systems equipped with HP 8200 PC user consoles that may impact patient safety. It was reported that images acquired with off-center collimation for extremities with the position selected as Digits to front or Digits to back are displayed as blank. Investigation into th
Source facts
- Action
- Consignees were sent a GE Healthcare "Product Safety Notification" letter dated March 7, 2008. The letter was addressed to Hospital Administrators, Mangers of Radiology and Radiologists. The letter listed the Safety Issue, Affected Product, Safety Instructions, Corrective Action and Contact Information.
- Initiated
- 2008-02-14
- Root cause
- Software design
- Terminated
- 2012-01-07
- Distribution
- Nationwide Distribution -- AL, AZ, CA, CO, FL, GA, HI, IL , IN, KY, LA, MD, MA, MI,MN,MS, MO, MT, NE, NH, NJ, NM, NY, NC, OH, OK OR, PA, SC, TN, TX, VA, WA, and WV.
- Recall number
- Z-0002-2009
- Recall status
- Terminated
- CDRH record ID
- 84800
- FDA classified
- 2009-09-24
- Recall event number
- 53157