Open original FDA record ↗Recall record
- FDA identifier
- Z-0002-2010
- Company in FDA record
- Edwards Lifesciences, Llc
- FDA classification date
- Nov 10, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Several complaints were the two parts of the tubular package became loose under the shrink wrap that holds the DFU on the packaging tube.
Source facts
- Action
- Edwards initiated a worldwide recall via customer letter dated June 29, 2009, sent via Federal express on same day, requesting that customers cease using Fogarty Occlusion Catheter, Model No. 62080822F products from the indicated lots and that they return all unused products. Edwards requested that the customers provide written verification of stock and identify any unused products to be returned to Edwards Lifesciences LLC, Att: Field Action Coordinator, ODW Logistics, Inc., 13725 Pipeline Ave., Chino, CA 91710. Additionally provided Customer Service Tel # 800.424.3278.
- Initiated
- 2009-06-22
- Root cause
- Packaging change control
- Terminated
- 2012-02-23
- Distribution
- AL AR AZ CA CO CT DC DE FL GA HI IA ID IL IN KS KY LA MA MD MI MN MO MS MO NC ND NE NH NJ NM NV NK OH OR PA RI SC SD TN TX UT VA VT WA WI WV WY, Puerto Rico Australia, Belgium, Brazil, Bulgaria, Canada, Canary Island, China, Czech Republic, Egypt, Finland, France, Germany, Great Britain, Greece, Hungary, Iceland, India, Ireland, Italy, Japan, Luxembourg, Netherland, New Zealand, Poland, Portugal, Rumania, Russia, Singapore, Slovak Republic, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, and Turkey.
- Recall number
- Z-0002-2010
- Recall status
- Terminated
- CDRH record ID
- 84106
- FDA classified
- 2009-11-10
- Recall event number
- 52834