Z-0002-2022: REF TIN3015, 15G x 79 mm Bone Marrow Aspiration/Intraosseous Infusion Needle, For Single Use Only, Steril

Recall record Z-0002-2022 for Bard Peripheral Vascular Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0002-2022
Company in FDA record
Bard Peripheral Vascular Inc
FDA record creation date
Oct 1, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for needle protector to become fully dislodge exposing infusion needle which could result in needle stick.

Source facts

Action
On August 4, 2021, Bard Peripheral Vascular, Inc. issued an "Urgent Medical device Correction" notice to affected customers via Fed Ex. In addition to informing consignees about the recall product, BD ask consignees to take the following actions: 1. Please check all inventory locations within your institution for the product listed in Attachment A and immediately destroy all affected product remaining in your possession. The product should have the sterile barrier broken and destroyed per your local facility s destruction process. 2. Share this notice with any users of the product within your facilities to ensure they are also aware of this Urgent Medical Device Correction. 3. If you purchased this product from a distributor, contact your distributor for further instructions. 4. Complete and return the attached Customer Response Form to the BD contact noted on the form confirming acknowledgement of the correction notification, whether or not you have any affected product, so that BD may acknowledge your receipt of this notification and process your credit. 5. Please contact your BD representative to assist in this process or you may also contact the number below for further assistance. 6. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program: Web: MedWatch website at www.fda.gov/medwatch Phone: 1-800-FDA-1088 (1-800-332-1088) Mail: MedWatch, HF-2, FDA, 5600 Fisher s Lane, Rockville, MD 20852-9787 7. BD Contact Information: If you require further assistance please contact: North American Regional Complaint Center, 1-844-8BD- LIFE (1-844-823-5433) Say Recall when prompted, M-F 8am - 5pm CT, Recall Questions, Product Complaints, Technical Questions
Initiated
2021-07-27
Root cause
Packaging change control
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, NC, NE, NH, NJ, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WV and the countries of Netherlands, United Arab Emirates, Australia, Taiwan, Asia, South Africa, Canada, JAPAN, India, Hong Kong, New Zealand, Mexico, China, Belgium, Malaysia, Singapore, Vietnam.
Recall number
Z-0002-2022
Recall status
Open, Classified
CDRH record ID
188744
FDA record created
2021-10-01
Recall event number
88473

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗