Skip to content

Public FDA record

Z-0002-2023: Ivenix Infusion System (IIS), LVP Software; LVP-SW-0004 version 5.2.0

Recall record Z-0002-2023 for Fresenius Kabi USA, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0002-2023
Company in FDA record
Fresenius Kabi USA, LLC
FDA classification date
Oct 6, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

The display screen may become frozen and unresponsive to user input, triggering an alarm that is evidenced by a buzzer sound and flashing red LEDs as a secondary notification. However, as the screen goes blank it does not indicate an alarm condition or respond. This issue leads to temporary cessation of flow of a medication or fluid, requiring immediate intervention by the clinician.

Source facts

Action
On September 8, 2022, the company notified affected customers via phone call and emailed letter. Customers were instructed to do the following: 1. If a high priority Pump Problem alarm occurs, the LVP should be powered down by holding the power button for 10 seconds then releasing the power button so that the screen goes dark, then pressing the power button again to re-start the pump. Alternatively, if you have another LVP available, you may use it to resume infusion. 2. Inform potential users of the product within your organization of this notification. If your facility further distributes or transfers products amongst satellite sites or other locations, please disseminate this information accordingly. Fresenius Kabi has updated the software to rectify the condition. The updated software is currently available. Please schedule a service call at your earliest convenience to have it installed. Customer support: Ivenix_support@fresenius-kabi.com; 1-978-775-8100
Initiated
2022-09-08
Root cause
Software design
Terminated
2024-09-10
Distribution
Domestic distribution to NJ and WI.
Recall number
Z-0002-2023
Recall status
Terminated
CDRH record ID
196036
FDA classified
2022-10-06
Recall event number
90845

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗