Z-0003-2007: Axis-Shield DIASTAT Anti-CCP test kit, (code FCCP200)

Recall record Z-0003-2007 for Axis-Shield Diagnostics, Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0003-2007
Company in FDA record
Axis-Shield Diagnostics, Ltd.
FDA classification date
Oct 3, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The preservative sodium azide used in the Kit Negative Control (part number FCOM175) was at the wrong concentration - the Kit Negative Control contains 0.2% sodium azide rather than the intended 0.1%.

Source facts

Action
Advisory letter was sent to customers informing them of the increased sodium azide content. A specific letter was sent to each customer / distributor informing them of the misformulation. The content of the letter referred to the specific kits received by each customer. The letter informed users about the additional risk relating to double azide concentration. The outcome of the risk evaluation was as follows: 1) There is no impact to product performance and hence patient results because of this. 2) The increase in the sodium azide concentration changes the hazard status from ''No significant hazard'' to ''Harmful'', with the following additional risk phrases R22 (harmful if swallowed) and R32 (contact with acids liberates toxic gas). The accompanying safety phrases are S23 (do not breathe fumes), S26 (in case of contact with eyes, rinse immediately with plenty of water and seek medical advice), S46 (if swallowed, seek medical advice immediately and show this container or label) and S61 (avoid release to environment, refer to special instructions/safety data sheets). 3) The risk of the increase in sodium azide concentration in the Kit Negative Conrol to the end user is deemed to be minimal due to the low volume of the Negative Control. The letter states that the product can still be safely used and users should be aware of and take notice of the advisory letter.
Initiated
2005-11-11
Root cause
Other
Terminated
2007-03-03
Distribution
Worldwide distribution ---- USA including states of MN, CA, NJ, WA, IL, CO,and NY and countries of Canada, Spain, Bulgaria, UK, Lebanon, Russia, Columbia, South Africa, Italy, Sweden, Germany, Norway, Venezuela, Japan, Tunisia, Cyprus, The Netherlands, Greece, Israel, and India.
Recall number
Z-0003-2007
Recall status
Terminated
CDRH record ID
47623
FDA classified
2006-10-03
Recall event number
36018

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗