Open original FDA record ↗Recall record
- FDA identifier
- Z-0003-2010
- Company in FDA record
- Philips Healthcare Inc.
- FDA classification date
- Nov 9, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
incorrect default configuration settings, insufficient battery power as the user may not be alerted that no battery is installed or that a low battery condition exists prior to disconnecting from AC or DC power
Source facts
- Action
- Philips Healthcare issued a Urgent Voluntary Medical Device Correction Notification on August 21, 2009 advising users of the problem and Customers are asked to follow the "Action to be taken by customer/user" section of the Urgent Voluntary Medical Device Correction Notification. In this section, customers are given the instructions to correct their devices. For further information or support concerning this issue, please contact Philips representative at 1-800-722-9377.
- Initiated
- 2009-08-21
- Root cause
- Process change control
- Terminated
- 2017-04-25
- Distribution
- Nationwide Canada Australia, Austria, Beharin, BRUNEI DARUSSALAM,Chian,. Finland, France, Germany,Hong Kong, India, Indonesia, Ireland, Israel,Italy, Japan, Libya, Lithuania. Mexico, Nambia, Netherlands, Palestine, Philippines, Poland, Portugal, Qatar, Russia, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, UAE, UK
- Recall number
- Z-0003-2010
- Recall status
- Terminated
- CDRH record ID
- 84415
- FDA classified
- 2009-11-09
- Recall event number
- 52990