Z-0004-2013: Male, 16Fr, Straight-tip Intermittent Urinary Catheters M16. The product is distributed individually or i

Recall record Z-0004-2013 for Cure Medical LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0004-2013
Company in FDA record
Cure Medical LLC
FDA classification date
Oct 2, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class II

Reason stated in the source

The firm initiated the product recall because labeling on some of the intermediate boxes of catheters identify the contents to be M16C, a coud-tip catheters while the actual contents are M16Fr male, 16Fr, straight-tip catheters with the pouch and the case box labeled as M16Fr male, 16Fr, straight-tip catheters.

Source facts

Action
Cure Medical LLC sent an Urgent Medical Device Correction and Removal Notification letter dated September 19, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. The letter instructed the customers to: a) Check their shelves for M16C with lot number 120425 and quarantine all affected product. b) Verify the contents (pouch label) are M16C and correspond with the intermediate box label. c) If all match, have the dealer sign and return a supplied verification record. d) If any mismatch is found, contact Cure Medical for assignment of a return goods authorization. e) Cure Medical will replace any incorrect product along with processing costs for this situation to the dealer. If affected product was further distributed customers were asked to identify those customers who received them and notify them of the problem. For questions customers were instructed to call 1-800-570-1778.
Initiated
2012-09-19
Root cause
Labeling mix-ups
Terminated
2013-02-07
Distribution
Nationwide Distribution
Recall number
Z-0004-2013
Recall status
Terminated
CDRH record ID
113089
FDA classified
2012-10-02
Recall event number
63241

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗