Open original FDA record ↗Recall record
- FDA identifier
- Z-0004-2017
- Company in FDA record
- Leonhard Lang Medizintechnik GmbH
- FDA classification date
- Oct 14, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
There is a risk that defibrillation electrodes model DF29N will be connected with the defibrillator Welch Allyn AED 10 only with delay or not at all. This may cause a situation in which a patient, who is in a life threatening condition and requires a defibrillation shock, cannot be treated in good time.
Source facts
- Action
- Consignees was sent via e-mail a Leonhard Lang "Important Safety Notice" dated September 1, 2016. The letter described the product being recalled, Description of the defect, Actions and time frame of the recall, & Compensation for the recalled electrodes. Advised consignees to inform all users within their organizations of the recall; and, secure and destroy the unused electrodes in their inventory. Confirmation of destruction can be done by completing and returning the "Confirmation of Destruction / Consumption" form by October 14, 2016. If the product was further distributed, they were to forward a copy of the Safety Notice. For questions contact sales staff.
- Initiated
- 2016-09-01
- Root cause
- Device Design
- Terminated
- 2019-03-26
- Distribution
- Worldwide Distribution - US, to the state of Florida; and, the countries of South Africa, Germany, Slovenia, France, United Arab Emirates, Great Britain, Italy, Israel, Serbia, Thailand, Thailand, South Africa, Lebanon, and Poland.
- Recall number
- Z-0004-2017
- Recall status
- Terminated
- CDRH record ID
- 149595
- FDA classified
- 2016-10-14
- Recall event number
- 75181