Open original FDA record ↗Recall record
- FDA identifier
- Z-0004-2023
- Company in FDA record
- Philips North America Llc
- FDA classification date
- Oct 4, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Device may intermittently fail to produce audio. Turning the unit off and on does not resolve the issue; disconnecting and reconnecting to AC power does not reliably resolve the issue.
Source facts
- Action
- On September 6, 2022, Philips notified affected customers through URGENT Medical Device Recall letters. A Philips representative will contact customers to arrange a software update. once arranged, a Philips representative will conduct a software update on the device for the customer. Customers should share the recall information with all clinical staff to review and understand. The notice should be placed with the documentation of the MR Patient Care Portal 5000.
- Initiated
- 2022-09-06
- Root cause
- Under Investigation by firm
- Distribution
- Domestic distribution to AL AZ CA CO FL GA ID IL IN LA MA MI MN NC NE NJ OH PA SC TN TX UT VT WA WY. Foreign distribution to Australia Austria Belgium Czech Republic Denmark Estonia Finland France Germany Ireland Italy Malta Netherlands Norway Poland Portugal Qatar Singapore Slovakia Spain Sweden Switzerland United Kingdom
- Recall number
- Z-0004-2023
- Recall status
- Open, Classified
- CDRH record ID
- 196075
- FDA classified
- 2022-10-04
- Recall event number
- 90857