Skip to content

Public FDA record

Z-0004-2023: MR Patient Care Portal 5000 (Desktop Unit), Model Number 453564792561 This is a component of product num

Recall record Z-0004-2023 for Philips North America Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0004-2023
Company in FDA record
Philips North America Llc
FDA classification date
Oct 4, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Device may intermittently fail to produce audio. Turning the unit off and on does not resolve the issue; disconnecting and reconnecting to AC power does not reliably resolve the issue.

Source facts

Action
On September 6, 2022, Philips notified affected customers through URGENT Medical Device Recall letters. A Philips representative will contact customers to arrange a software update. once arranged, a Philips representative will conduct a software update on the device for the customer. Customers should share the recall information with all clinical staff to review and understand. The notice should be placed with the documentation of the MR Patient Care Portal 5000.
Initiated
2022-09-06
Root cause
Under Investigation by firm
Distribution
Domestic distribution to AL AZ CA CO FL GA ID IL IN LA MA MI MN NC NE NJ OH PA SC TN TX UT VT WA WY. Foreign distribution to Australia Austria Belgium Czech Republic Denmark Estonia Finland France Germany Ireland Italy Malta Netherlands Norway Poland Portugal Qatar Singapore Slovakia Spain Sweden Switzerland United Kingdom
Recall number
Z-0004-2023
Recall status
Open, Classified
CDRH record ID
196075
FDA classified
2022-10-04
Recall event number
90857

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗