Some hard shell carry case for LIFEPAK 500 and LIFEPAK CR Plus automated external defibrillators may have blocked pressure vent which may make it difficult to open the case after a sudden decrease in altitude (such as in aircraft landng or mountain descent).
Source facts
Action
On 7/12/05 the firm issued a letter to customers instructing the customers to inspect the cases for blocked vent. If the customer noted that the vent was blocked, to contact the firm for a replacement case. The firm provided response cards for the customers to return stating they had inspected their units.
Initiated
2005-07-12
Root cause
Process control
Terminated
2006-06-26
Distribution
Cases were distributed nationwide and internationally.
Recall number
Z-0005-06
Recall status
Terminated
CDRH record ID
40483
FDA classified
2005-10-04
Recall event number
32642