Audible signal is not available at all times when using the HLC fluoroscopy mode. Special means of activation are not designed in for selection of High Level Controls, as required by regulation.
Source facts
Action
Toshiba issued a letter entitled Important Electronic Product Radiation Warning, September 30, 2008, explaining the issue. The customers were requested to reply to the letter via fax-back form, and a Toshiba service representative would install the corrective action.
Initiated
2008-09-29
Root cause
Software design
Terminated
2012-10-04
Distribution
Nationwide
Recall number
Z-0005-2009
Recall status
Terminated
CDRH record ID
85541
FDA classified
2009-09-28
Recall event number
53451