Z-0005-2016: Celesteion PCA-9000A/2 PET/CT System

Recall record Z-0005-2016 for Toshiba American Medical Systems Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0005-2016
Company in FDA record
Toshiba American Medical Systems Inc
FDA classification date
Oct 7, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

It was found that if specific operations are performed in multi-phase helical scanning, the acquired raw data may not be saved due to a software problem.

Source facts

Action
Planned Action by Toshiba Medical America Systems to address the defect: a. A notification will be sent to the customer with affected system in compliance with 21 CFR1 003.21 upon approval by the Secretary. b. To correct this issue revised software will be installed on the system to prevent this occurrence from happening. The consignee will be contacted by their Toshiba service representative for an appointment when the revised software becomes available. c. The customer will be advised to do the following until the corrective measures have been taken, i.e., revised software installed on their systems. If you have any questions please contact your local Toshiba Representative at (800) 521-1968.
Initiated
2015-05-08
Root cause
Device Design
Terminated
2015-12-03
Distribution
US Distribution to the state of Nevada.
Recall number
Z-0005-2016
Recall status
Terminated
CDRH record ID
138996
FDA classified
2015-10-07
Recall event number
71763

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗