Open original FDA record ↗Recall record
- FDA identifier
- Z-0006-2008
- Company in FDA record
- Datex - Ohmeda, Inc
- FDA classification date
- Oct 4, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Elevated inspired CO2 levels related to the use of the Datex-Ohmeda GE Healthcare Disposable Multi Absorber Canister with the EZchange Module.
Source facts
- Action
- Because this issue only occurs when the Disposable Multi Absorber is used in conjunction with EZchange Module, they have contacted all EZchange Module customers. It is critical to contact these customers specifically, rather than all customers shipped Disposable Multi Absorbers as this may cause unwarranted confusion with customers who do not have the EZchange Module. Consignees were sent an letter dated 7/19/2007 describing the product and problem. The letter lists a number of interim actions to implement with the anesthesia systems.
- Initiated
- 2007-07-24
- Root cause
- Vendor change control
- Terminated
- 2015-10-16
- Distribution
- Worldwide, including USA, Australia, Belgium, Brazil, Canada, Chile, China, Denmark, Egypt, Estonia, Finland, France, Germany, India, Ireland, Italy, Jordan, Republic of Korea, Kuwait, Mexico, Netherlands, New Zealand, Norway, Peru, Poland, Qatar, Singapore, Slovak Republic, South Africa, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom, and Venezuela.
- Recall number
- Z-0006-2008
- Recall status
- Terminated
- CDRH record ID
- 53847
- FDA classified
- 2007-10-04
- Recall event number
- 38484