Open original FDA record ↗Recall record
- FDA identifier
- Z-0006-2009
- Company in FDA record
- Roche Diagnostics Corp.
- FDA classification date
- Oct 9, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Some of the labels are glossy, resulting in the bar code being difficult to read by the bar code reader.
Source facts
- Action
- Consignees were notified via an Urgent Medical Device correction letter dated 6/11/08 informing users of the problem and providing them with work-around instructions. The instructions suggests 1) inserting the cassette rack slowly; 2) using mainly reagent slots A, B, or C; 3) tilting the rack slightly clockwise so the barcode scanner does not read the barcode at an exact 90 degree angle; and 4) confirming the rack is properly seated on the track and displayed on the Cassette Status screen. If the cassette barcode does not register after following the Work-Around instructions, then customers are required to contact Diagnostics Technical Support.
- Initiated
- 2008-06-11
- Root cause
- Labeling Change Control
- Terminated
- 2009-09-22
- Distribution
- Nationwide Distribution
- Recall number
- Z-0006-2009
- Recall status
- Terminated
- CDRH record ID
- 71772
- FDA classified
- 2008-10-09
- Recall event number
- 48638