Skip to content

Public FDA record

Z-0006-2025: Miami J Select Collar, Catalog: MJS-101; Miami J Select Collar Set, Catalog: MJSR-101

Recall record Z-0006-2025 for Ossur H / F. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0006-2025
Company in FDA record
Ossur H / F
FDA classification date
Oct 1, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Pressure sensitive adhesive hooks attached to the sides of the anterior cervical orthosis collar panel can occasionally exhibit poor adhesion, which may lead to them gradually peeling off with repeated doffing, possibly leading to reduced immobilization of the cervical spine.

Source facts

Action
On 8/20/24, recall notices were mailed and emailed to customers who were asked to do the following: 1) Pass this notice to those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. 2) Maintain awareness on this notice for an appropriate period. 3) If you have further distributed this device, please identify your customers, and notify them at once of this recall. We recommend that you include a copy of this recall letter. 4) Firm recommends healthcare professionals perform a detailed inspection of the device to verify lot number, and to ensure structural stability for any patients currently wearing impacted device. 5) If any of your customers are currently wearing a device, the packaging label and product label should be inspected for the lot number: -If the lot number is outside of the affected range, no actions need to be taken. -If the lot number is inside the affected range or if the packaging label or lot number on the product label is not available, healthcare professionals are advised to contact customer service for replacement device. 6) Complete and return the acknowledgement form via email to acknowledgements@ossur.com Contact customer service at this phone number for further information and assistance: Tel: 800-233-6263
Initiated
2024-08-20
Root cause
Nonconforming Material/Component
Distribution
US: NH, MA, NY, NJ, PA, GA, MD, NC, IN, FL, AR, TN, KY, OH, MI, IL, TX, MO, CO, UT, AZ, WA, OR, CA, NM, VA, IA, MN, NE, SC, MT, LA, ID, SD, AL, CT, OK, AK, PR
Recall number
Z-0006-2025
Recall status
Open, Classified
CDRH record ID
209890
FDA classified
2024-10-01
Recall event number
95278

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗