Reason stated in the source
It was discovered that the operator console did not have a label affixed in accordance with 21CFR 1010.2 indicating certification to applicable performance standards.
Source facts
- Action
- GE Healthcare will bring customer systems into compliance with 21 CFR by having a GE Healthcare field service representative, who will perform a field corrective action, visit each customer site. This will be carried out via a Field Modification Instruction planned to be released in October 2009 and planned to be completed by February 2010. Additionally, all forward production operator consoles are manufactured so as to comply with 21 CFR and to include a manufacturing check for appropriate labeling.
- Initiated
- 2009-05-01
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2012-05-16
- Distribution
- Worldwide Distribution -- USA, including states of ME, FL, IA, AL, CT, CA, TX, KS, KY, OH, and WV and countries of Japan, Russian Federation, India, Argentina, Brazil, Italy, Poland, Puerto Rico, Romania, and Saudi Arabia.
- Recall number
- Z-0007-2011
- Recall status
- Terminated
- CDRH record ID
- 87691
- FDA classified
- 2010-12-17
- Recall event number
- 54154