Open original FDA record ↗Recall record
- FDA identifier
- Z-0007-2025
- Company in FDA record
- Philips North America
- FDA classification date
- Oct 1, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The 866173 lntelliVue G7m Anesthesia Gas Modules could have a component issue that may trigger an interruption of the gas measurement. This issue concerns the G7m's sample gas pump which could stop working, producing a technical, INOP (inoperable) alarm. The pump may fail prematurely when there is friction caused by an abrasion. This recall is for an IMPORTANT PRODUCT NOTICE dated 1/18/23.
Source facts
- Action
- An IMPORTANT PRODUCT NOTICE dated 1/18/23 was sent to customers. The actions planned by Philips to correct the problem A Philips representative will reach out to you to arrange a replacement of the sample gas pump. Affected products and how to identify them Affected product is the 866173 lntelliVue G7m Anesthesia Gas Module. Both the product number and the serial number on the G7m product is contained on the back of the product, preceded by "REF" and "SN", respectively. If you need any further information or support concerning this issue, please contact your local Philips representative.
- Initiated
- 2023-01-18
- Root cause
- Under Investigation by firm
- Distribution
- US Nationwide. Global Distribution.
- Recall number
- Z-0007-2025
- Recall status
- Open, Classified
- CDRH record ID
- 209939
- FDA classified
- 2024-10-01
- Recall event number
- 95301