The wrong electrical connector pin was attached to the cord during assembly by the manufacturer. this electrical connector is too big to fit into the port on the foot activated generator. this makes the device unusable.
Source facts
Action
Consignees were first notified by phone with a follow up e-mail or fax. The consignees were asked to complete the response section regarding the destruction/return of the device.
For questions regarding this recall call 425-882-1179.
Initiated
2012-09-06
Root cause
Nonconforming Material/Component
Terminated
2012-10-22
Distribution
Nationwide Distribution
Recall number
Z-0008-2013
Recall status
Terminated
CDRH record ID
113109
FDA classified
2012-10-03
Recall event number
63253