Open original FDA record ↗Recall record
- FDA identifier
- Z-0008-2015
- Company in FDA record
- Respironics California Inc
- FDA classification date
- Oct 21, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
A failure of a specific component may prevent the ventilators from operating on AC power or transitioning back to AC power when operating on battery. Additionally, if a battery is not present or is depleted the ventilators will not operate. This failure will cause the loss of ventilator support, which may result in hypercarbia or hypoxemia.
Source facts
- Action
- Philips Healthcare sent an Urgent - Field Safety Notice dated September 17, 2014, to all customers who purchased the Esprit V1000 and V200 Ventilators to inform them that Respironics is recalling Esprit V1000 and V200 Ventilators installed with 3rd Generation Power Supplies and also 3rd Generation Power Supply Repair Part Kits. The letter identifies the affected devices, informs the customers of the problems, hazard involved, and the actions to be taken. Customers were reminded to be advised of the warnings provided. A Philips Respironics representative will contact customers to schedule a no-cost replacement for further information or support. Customers with questions were instructed to contact Philips Healthcare at 1-800-722-9377. For questions regarding this recall call 760-918-7300.
- Initiated
- 2014-09-17
- Root cause
- Unknown/Undetermined by firm
- Terminated
- 2016-09-28
- Distribution
- Worldwide Distribution - USA, Argentina, China, Egypt, Ethiopia, Honduras, Hong Kong, India, Indonesia, Iraq, Israel, Italy, Jordan, Libyan Arab Yamahiriya, Malaysia, Mali, Mexico, Myanmar (Burma). Nepal, Nigeria. Pakistan, Peru, Philippines, Serbia, Singapore, South Africa, Sri Lanka, United Republic of Tanzania, Thailand, Vietnam, Zimbabwe, Republic of Korea, and Netherlands.
- Recall number
- Z-0008-2015
- Recall status
- Terminated
- CDRH record ID
- 130205
- FDA classified
- 2014-10-21
- Recall event number
- 69313