Skip to content

Public FDA record

Z-0009-2009: Roche Diagnostics STFR Tina-quant Soluble Transferrin Receptor, COBAS INTEGRA, cobas c systems, Roche Dia

Recall record Z-0009-2009 for Roche Diagnostics Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0009-2009
Company in FDA record
Roche Diagnostics Corp.
FDA classification date
Oct 9, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Some of the labels are glossy, resulting in the bar code being difficult to read by the bar code reader.

Source facts

Action
Consignees were notified via an Urgent Medical Device correction letter dated 6/11/08 informing users of the problem and providing them with work-around instructions. The instructions suggests 1) inserting the cassette rack slowly; 2) using mainly reagent slots A, B, or C; 3) tilting the rack slightly clockwise so the barcode scanner does not read the barcode at an exact 90 degree angle; and 4) confirming the rack is properly seated on the track and displayed on the Cassette Status screen. If the cassette barcode does not register after following the Work-Around instructions, then customers are required to contact Diagnostics Technical Support.
Initiated
2008-06-11
Root cause
Labeling Change Control
Terminated
2009-09-22
Distribution
Nationwide Distribution
Recall number
Z-0009-2009
Recall status
Terminated
CDRH record ID
71819
FDA classified
2008-10-09
Recall event number
48638

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗