Several complaints of pumps exhibiting A47 alarms were received. Investigation revealed the alarms occurred when the pump attempted to display text that exceeded the maximum display length of the pump screen. The investigation revealed this only occurs when the pump is programmed to display information in Spanish.
Source facts
Initiated
2004-09-14
Root cause
Other
Terminated
2005-03-03
Distribution
Nationwide & Spain
Recall number
Z-0010-05
Recall status
Terminated
CDRH record ID
35097
FDA record created
2015-06-25
Recall event number
30059