Skip to content

Public FDA record

Z-0010-05: Medtronic MiniMed Paradigm Model 712

Recall record Z-0010-05 for Medtronic MiniMed. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0010-05
Company in FDA record
Medtronic MiniMed
FDA record creation date
Jun 25, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Several complaints of pumps exhibiting A47 alarms were received. Investigation revealed the alarms occurred when the pump attempted to display text that exceeded the maximum display length of the pump screen. The investigation revealed this only occurs when the pump is programmed to display information in Spanish.

Source facts

Initiated
2004-09-14
Root cause
Other
Terminated
2005-03-03
Distribution
Nationwide & Spain
Recall number
Z-0010-05
Recall status
Terminated
CDRH record ID
35097
FDA record created
2015-06-25
Recall event number
30059

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 21, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗