Incorrect results, without a data flag, reported if one of the other reagent containers becomes depleted before the R1 reagent is depleted.
Source facts
Action
Customers were notified by certified mail on or about May 12, 2003. Operators were given a workaround (assure R1 has the lower number of tests remaining before using the analyzer) until new software is supplied.
Initiated
2003-05-08
Root cause
Other
Terminated
2003-12-09
Distribution
United States
Recall number
Z-0012-04
Recall status
Terminated
CDRH record ID
27322
FDA classified
2003-10-15
Recall event number
26231