Z-0012-04: Roche/Hitachi Modular Analytics P clinical chemistry analyzer; Modular P. Catalog number 766-8156.

Recall record Z-0012-04 for Roche Diagnostics Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0012-04
Company in FDA record
Roche Diagnostics Corp.
FDA classification date
Oct 15, 2003
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Incorrect results, without a data flag, reported if one of the other reagent containers becomes depleted before the R1 reagent is depleted.

Source facts

Action
Customers were notified by certified mail on or about May 12, 2003. Operators were given a workaround (assure R1 has the lower number of tests remaining before using the analyzer) until new software is supplied.
Initiated
2003-05-08
Root cause
Other
Terminated
2003-12-09
Distribution
United States
Recall number
Z-0012-04
Recall status
Terminated
CDRH record ID
27322
FDA classified
2003-10-15
Recall event number
26231

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗