Upon evaluation, the firm found that the fatigue strength of these nails was below the required value as specified in the design specification.
Source facts
Action
The firm sent out via e-mail notification to their field representatives on 6/27/2005 to identify, remove and return to EBI all affected product from their inventories. One direct customer, Dr. Maylack was sent a recall letter on 7/7/2005.
Initiated
2005-06-24
Root cause
Other
Terminated
2006-07-03
Distribution
There are 9 sales representatives that are instructed to remove and return the EBI nail from nationwide customers.
Recall number
Z-0012-06
Recall status
Terminated
CDRH record ID
40870
FDA classified
2005-10-07
Recall event number
32813