Philips Medical Systems North America Co. Phillips
FDA classification date
Oct 6, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification
Reason stated in the source
When a patient is selected in the patient list and examination tab pressed with background query running at the same time, the wrong patient may be selected.
Source facts
Action
Philip Medical System sent an "URGENT-Device Correction" letter dated August 13, 2008 to their customers. The letter describes the recalled product as Essenta DR Patient and Image Data Software (SW) release 1.0. The firm informs the customers that when patient is selected in patient list and examination tab pressed with background query running at the same time, there is a risk that the wrong patient is selected on the Essenta DR Computed Radiography System.
Customers are advised to verify the selected patient's information before releasing an exposure while the firm works on upgrading the Software. All affected units will receive a software upgrade to correct the issue.
Customers should contact Philips Customer Care Service at 800-772-9377, #5, #4, #1 and reference FCO 712000031 for any questions or support concerning the recall.
Initiated
2008-08-15
Root cause
Other
Terminated
2010-10-07
Distribution
Nationwide distribution: CA, GA, FL, KS, MN, NY, OH, OR, SC, TN, TX, UT, and WA.
Recall number
Z-0012-2011
Recall status
Terminated
CDRH record ID
93207
FDA classified
2010-10-06
Recall event number
49138