Open original FDA record ↗Recall record
- FDA identifier
- Z-0012-2018
- Company in FDA record
- Toshiba American Medical Systems Inc
- FDA record creation date
- Oct 17, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
During an examination images were displayed on the live monitor,but the images were not displayed on the system monitor nor were they saved to the hard disk.
Source facts
- Action
- Toshiba planned action to bring defect into compliance: 1. You will issue a Customer Notification Letter to make customers aware of the hazards, what measures will be taken to remedy the defect at no cost to the user, and how to safely use the equipment before corrections can be made. 2. You will install modified firmware on the systems to correct the defect and prevent recurrence. 3. You will provide progress reports on the effectiveness of the plan. CDRH approves the CAP subject to the following conditions: 1. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator are to be included in the notification. Thank you for your urgent attention to this matter. If you have any questions regarding this letter please feel free to call (800) 421-1968, or contact your local Toshiba Representative at (800) 521-1968. For further questions please call (714) 730-5000.
- Initiated
- 2017-09-12
- Root cause
- Software design
- Terminated
- 2020-11-18
- Distribution
- USA (nationwide) Distribution
- Recall number
- Z-0012-2018
- Recall status
- Terminated
- CDRH record ID
- 158813
- FDA record created
- 2017-10-17
- Recall event number
- 78166