Docking screw/nut failure, which may result in a treatment head dropping, potentially impacting the patient.
Source facts
Action
Recall initiated on 06/20/2008, advising users via Advisory Notice that a safety upgrade would be available. Two options are given to either self inspect the product or schedule an Intraop Authorized Field Service Engineer to visit their facility to perform the inspection. The letter notes that if no nut failure was evident, the machine could still be used. The firm continues to send retrofit kits.
Initiated
2008-06-20
Root cause
Component design/selection
Terminated
2011-04-05
Distribution
Worldwide Distribution
Recall number
Z-0014-2009
Recall status
Terminated
CDRH record ID
72537
FDA classified
2008-10-03
Recall event number
49022