Open original FDA record ↗Recall record
- FDA identifier
- Z-0014-2013
- Company in FDA record
- Baxter Healthcare Corp.
- FDA classification date
- Oct 9, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class I
Reason stated in the source
Baxter has determined that the ball-valve feature of the Buretrol Solution Sets may not function as expected, allowing air to flow past the valve and into the tubing at the completion of dose within the burette.
Source facts
- Action
- Baxter Healthcare Corp. sent an Urgent Product Recall letter dated September 7, 2012, via first class mail to all affected customers, informing them of the problem with the ball-valve, and instructing them to discontinue use of the affected product and to contact Baxter for instructions to return the affected product.. The customers were also requested to complete the attached customer reply form, confirming their receipt of the letter, and fax it to Baxter at the number provided on the form. Any questions regarding the communication were directed to Medical Information Services at Baxter at 1-800-933-0303.
- Initiated
- 2012-09-07
- Root cause
- Device Design
- Terminated
- 2017-12-29
- Distribution
- Worldwide Distribution - US Nationwide and Internationally to Canada, Costa Rica, Australia, New Zealand, Hong Kong, and the United Kingdom.
- Recall number
- Z-0014-2013
- Recall status
- Terminated
- CDRH record ID
- 112403
- FDA classified
- 2012-10-09
- Recall event number
- 63059