A limited amount of Carina mobile lifts have been delivered with a too small locking pin. If a locking pin
smaller than the correct size is mounted on the Carina mobile lift, the locking pin is likely to cause
increased wear on the plastic bushings. In case of continued use, these components may cause wear on the metal of the piston rod eye.
Source facts
Action
The firm initiated their recall by email on 07/28/2017. The firm directed the consignee to repair the units on site and destroy the affected parts.
Initiated
2016-07-28
Root cause
Process control
Terminated
2018-12-17
Distribution
OH, PA, MD
Recall number
Z-0014-2018
Recall status
Terminated
CDRH record ID
158744
FDA record created
2017-10-11
Recall event number
78130