Mouse anti-TTF-1 had become contaminated by immunoglobulins from mouse serum that was co-eluted during purification leading to false readings.
Source facts
Action
The firm has conducted telephone calls and visits to its consignees, with confirmation done by phone and fax. The firm has recovered or accounted for 100% of the products, either returned or destroyed.
Initiated
2004-04-06
Root cause
Other
Terminated
2004-10-29
Distribution
The Firm distributed to a total of 8 domestic consignees and an unspecified number of foreign consignees. The products were distributed to MD, GA, TX, OH, IL, DC, NY and CA. Foreign countries distributed to include France, Portugal, China, Italy, Israel, Philippines, and Germany.
Recall number
Z-0015-05
Recall status
Terminated
CDRH record ID
34912
FDA classified
2004-10-09
Recall event number
29984
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.