Method is susceptible to positive interference from endogenous ALT in patient samples.
Source facts
Action
The firm has updated the Limitations section of the Ammonia Reagent Package Insert and circulated a Technical Bulletin dated July 24, 2007 to all customers that have purchased Ammonia Reagent that is still within expiration date. A Notification of Change has been introduced and shall be placed within subsequent Ammonia Reagent Kits with a copy of the updated package insert; the notification of change informs customers that a change has been made to the package insert. Notifications of Change are placed within kits for 6 months from the date of issue.
Initiated
2007-07-24
Root cause
Device Design
Terminated
2007-12-18
Distribution
Worldwide, including USA, Ecuador, and Canada.
Recall number
Z-0015-2008
Recall status
Terminated
CDRH record ID
55366
FDA classified
2007-10-04
Recall event number
39389