The affected lots may be packaged in syringes which include an incorrect bung. The incorrect bung is made of a latex-free isoprene rubber material. After prolonged storage and exposure to contained solutions, the rubber bung may degrade due to chemical interaction with the bleach solution.
Source facts
Action
The firm initiated the recall by letter on 06/27/2019. The letter directed the consignee to quarantine the product and contact the returns department for return instructions.
Initiated
2019-06-27
Root cause
Process control
Terminated
2021-08-10
Distribution
US, South Africa, Switzerland, Poland, Italy, Singapore, Australia
Recall number
Z-0016-2020
Recall status
Terminated
CDRH record ID
175287
FDA classified
2019-10-01
Recall event number
83435
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.