Skip to content

Public FDA record

Z-0016-2022: Scan Abutment SU, Cat. No. 30-70104 - Product Usage: used to take an impression with an intra oral scann

Recall record Z-0016-2022 for Paltop Advanced Dental Solutions Ltd. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0016-2022
Company in FDA record
Paltop Advanced Dental Solutions Ltd
FDA record creation date
Oct 4, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Some devices are outside tolerances and showed a maximal deviation of 2 degrees in the orientation between the peek head and titanium base, and an incongruence between the size engraved on the peek has been identified. Deviations may delay the final restoration.

Source facts

Action
Impacted customers were contacted on July 26, 2021 via overnight delivery of written notification. The firm is replacing all components listed in the recall, purchased prior to 7/21/2021. Customers were asked to complete and return the provided product replacement form. A customer service representative will contact you to evaluate your inventory and to facilitate the exchange immediately. Please contact your Customer Service Department at 866-902-9272 for any questions.
Initiated
2021-07-26
Root cause
Process change control
Terminated
2023-09-21
Distribution
US Nationwide distribution. No foreign distribution from US.
Recall number
Z-0016-2022
Recall status
Terminated
CDRH record ID
189022
FDA record created
2021-10-04
Recall event number
88576

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗