Open original FDA record ↗Recall record
- FDA identifier
- Z-0016-2026
- Company in FDA record
- Getinge Disinfection Ab
- FDA classification date
- Oct 1, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.
Source facts
- Action
- Getinge notified consignees on about 08/29/2025 via letter. Consignees were instructed that Getinge service technicians will be proactively reaching out to ensure affected devices are calibrated appropriately. In the meantime, Getinge has developed a verification procedure to ensure that the circulation pressure in standby mode is correctly calibrated to 0.0 kPa. If the value deviates, recalibration by a qualified technician is required. The verification of the circulation pressure shall be performed by the operator or Getinge technician. If the value deviates from 0.0 kPa out of allowed tolerances (1 kPa), stop using the device immediately and work with your Getinge service technician to arrange recalibration. Consignees were also instructed to ensure all caregivers and users of the affected product are made aware of the recall and to complete and return the provided response form. ***UPDATE 10/16/2025: Firm notified FDA of an expansion to include additional OUS units and one additional complaint. ***UPDATE 03/02/2026: Firm notified FDA of an expansion to include additional OUS units and one additional complaint.
- Initiated
- 2025-08-29
- Root cause
- Under Investigation by firm
- Distribution
- Worldwide - US Nationwide distribution in the states of Colorado, Massachusetts, New Mexico, New Hampshire and the countries of Algeria, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Austria, Hong Kong, Hungary, India, Iraq, Ireland, Israel, Italy, Japan, Kenya, Korea, Republic of, Kuwait, Lithuania, Mexico, Netherlands, Norway, New Zealand, Poland, Qatar, Romania, Saudi Arabia, Singapore, Spain, Sweden, Thailand, Turkey, United Arab Emirates, United Kingdom.
- Recall number
- Z-0016-2026
- Recall status
- Open, Classified
- CDRH record ID
- 215875
- FDA classified
- 2025-10-01
- Recall event number
- 97535