Z-0016-2026: Getinge 88-Series Washer- Disinfector. Model Number: 88-5.

Recall record Z-0016-2026 for Getinge Disinfection Ab. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0016-2026
Company in FDA record
Getinge Disinfection Ab
FDA classification date
Oct 1, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.

Source facts

Action
Getinge notified consignees on about 08/29/2025 via letter. Consignees were instructed that Getinge service technicians will be proactively reaching out to ensure affected devices are calibrated appropriately. In the meantime, Getinge has developed a verification procedure to ensure that the circulation pressure in standby mode is correctly calibrated to 0.0 kPa. If the value deviates, recalibration by a qualified technician is required. The verification of the circulation pressure shall be performed by the operator or Getinge technician. If the value deviates from 0.0 kPa out of allowed tolerances (1 kPa), stop using the device immediately and work with your Getinge service technician to arrange recalibration. Consignees were also instructed to ensure all caregivers and users of the affected product are made aware of the recall and to complete and return the provided response form. ***UPDATE 10/16/2025: Firm notified FDA of an expansion to include additional OUS units and one additional complaint. ***UPDATE 03/02/2026: Firm notified FDA of an expansion to include additional OUS units and one additional complaint.
Initiated
2025-08-29
Root cause
Under Investigation by firm
Distribution
Worldwide - US Nationwide distribution in the states of Colorado, Massachusetts, New Mexico, New Hampshire and the countries of Algeria, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Austria, Hong Kong, Hungary, India, Iraq, Ireland, Israel, Italy, Japan, Kenya, Korea, Republic of, Kuwait, Lithuania, Mexico, Netherlands, Norway, New Zealand, Poland, Qatar, Romania, Saudi Arabia, Singapore, Spain, Sweden, Thailand, Turkey, United Arab Emirates, United Kingdom.
Recall number
Z-0016-2026
Recall status
Open, Classified
CDRH record ID
215875
FDA classified
2025-10-01
Recall event number
97535

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗