Open original FDA record ↗Recall record
- FDA identifier
- Z-0018-2011
- Company in FDA record
- Draeger Medical Systems, Inc.
- FDA classification date
- Oct 7, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Patients received insufficient ventilation. The Oxylog 3000 Instructions are inadequate relating to dead space volume.
Source facts
- Action
- The firm sent Important Safety Notice Letter, dated August 2010, and Instructions for Use supplement to consignees. The letter explained the issue with the affected product. The letter asks customers to attach the Instructions for Use amendment sheet supplied with the letter to their corresponding Instructions for Use. The firm also states that a they will provide a dedicated pediatric breathing circuit for the Oxylog 3000. Questions regarding the letter should be directed to Michael Kelhart at 1-800-543-5047. Questions regarding the use of the device should be directed to DragerService Technical Support at 1-800-543-5047 or your local Drager Sales Representative.
- Initiated
- 2010-08-19
- Root cause
- Error in labeling
- Terminated
- 2010-11-18
- Distribution
- Nationwide Distribution
- Recall number
- Z-0018-2011
- Recall status
- Terminated
- CDRH record ID
- 94060
- FDA classified
- 2010-10-07
- Recall event number
- 56600