Product was assembled with an isolating fiberglass hose within the oven in the wrong position. This results in improper shielding from potential electrical shock.
Source facts
Action
The firm issued a Field Action Notice by letter on 09/05/2018. The notice explained the issue and the possible hazard. The firm will schedule a visit with a service engineer to perform the correction.
Initiated
2018-09-04
Root cause
Process design
Terminated
2020-07-10
Distribution
Distribution in US state of MO.
Recall number
Z-0019-2019
Recall status
Terminated
CDRH record ID
167177
FDA record created
2018-10-03
Recall event number
80986