Z-0020-2025: IntelliVue Patient Monitor MX400/MX450/MX500/MX550 with software version P.01.01 and ordered with Option

Recall record Z-0020-2025 for Philips North America Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0020-2025
Company in FDA record
Philips North America Llc
FDA classification date
Oct 4, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Devices with Option CP2 and initially programmed to software version P.01.01 were configured incorrectly at the factory. Therefore, Enhanced ECG Capabilities provided by Option CP2 were not enabled in these devices. Without Option C01 Full Arrhythmia, the device will not provide the yellow alarms for enhanced arrhythmia detection. This is a retrospectively reported recall from 6/14/23.

Source facts

Action
PHILIPS issued Urgent Field Safety Notice (2023-CC-HPM-019) on 6/14/23. Letter states reason for recall, health risk and action to take: " Please be aware that without Option C01 (Full Arrhythmia) the device will not provide the yellow alarms for enhanced arrhythmia detection. " Review the contents of this letter with your staff. " Pass this notice to all those who need to be aware within your organization or to any organization where the affected devices might have been transferred. 5. Actions taken by Philips to correct the problem Philips Representative will contact you to arrange reload of current device's software to enable missing software options (C0l, C13, C51, C54). If you need any further information, please contact your local Philips representative
Initiated
2023-06-14
Root cause
Process control
Distribution
Foreign Only: Denmark France Germany Japan United Kingdom
Recall number
Z-0020-2025
Recall status
Open, Classified
CDRH record ID
209981
FDA classified
2024-10-04
Recall event number
95324

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗