The tracker is not locking adequately to the anchoring pin, resulting in unacceptable movement, which will affected the patient tracker system.
Source facts
Action
Sales force was notified these pre-launch devices are not to be used for surgical use via e-mail and telephone calls on 7/14/04.
Initiated
2004-07-14
Root cause
Other
Terminated
2004-10-18
Distribution
United States.
Recall number
Z-0021-05
Recall status
Terminated
CDRH record ID
34394
FDA classified
2004-10-19
Recall event number
29675