Vaporizer shutdown with alarm-An intermittent leak in the Aladin 2 cassette/Aisys pneumatic interface (valves) may occur that will invoke an alarm state. When this occurs 'Vaporizer Failure' will appear on the Aisys main display, with an accompanying audible alrm. Delivery of anesthetic Agent is halted; gas flow and ventilation are not affected.
Source facts
Action
Firm mailed recall letters on August 10, 2006, to consignees telling them what to do if a failure occurred during system checkout or during use, as a short term solution. Later, GE representatives will make visits to each site for modification of the interface valve on each device, as the long term solution.
Initiated
2006-08-10
Root cause
Other
Terminated
2008-02-24
Distribution
AK, AL, CA, FL, KS, KY, MD, MN, MO, NC, NE, SC, SD, TX, VA, WI,
Recall number
Z-0021-2007
Recall status
Terminated
CDRH record ID
47879
FDA classified
2006-11-07
Recall event number
36057