Skip to content

Public FDA record

Z-0021-2025: The Trevo Trak 21 Microcatheter is a single-lumen, braided shaft, variable stiffness catheter with radiop

Recall record Z-0021-2025 for Stryker Neurovascular. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0021-2025
Company in FDA record
Stryker Neurovascular
FDA classification date
Oct 7, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Stryker Neurovascular has observed that devices with Out of United States (OUS) approved Instructions for Use (IFU) and Packaging Labels (pouch and carton), different configurations, have been erroneously distributed to customers in the US.

Source facts

Action
On 09/12/2024, the firm sent via FedEx an "URGENT: MEDICAL DEVICE RECALL - Customer Notification" Letter to customers informing them that certain lots of Trevo Trak 21 Microcatheters were shipped with the incorrect configuration. The affected products contained Instructions for Use (IFUs) and product labels for international use, instead of the United States (US). The International IFUs contain intended uses that have not been cleared/approved for the U.S. market. Customer are instructed to: 1. Immediately check internal inventory for affected devices. 2. When using impacted lot(s) of Trevo Trak 21 Microcatheters, follow the US IFU attached in this communication NV00055463-02 (Attachment A) instead of the one that originally came with the device. 3. Circulate recall notification internally to all interested/affected parties and post this customer notification on or near affected products. 4. Maintain awareness of this communication internally until all required actions have been completed within your facility. 5. Inform Stryker Neurovascular if the subject devices have been distributed to other organizations and provide contact details so that Stryker Neurovascular can inform the recipients appropriately. 6. Inform Stryker Neurovascular of any adverse events concerning the use of the subject devices. (US Customer Service: 1-855-91-NEURO (1-855-916- 3876), Email: NVCustomerCare@stryker.com). 7. Complete the attached customer response form. It may be that you no longer have any physical inventory on site. Completing this form allow for updating records and will also negate the need to send any further unnecessary communications on this matter. 8. Email the completed form to nvfieldactions@stryker.com. Contact the local Stryker representative, send an email to nvfieldactions@stryker.com or call US Customer Service: 1-855-91-NEURO (1-855-916-3876) for further assistance related to this Medical Device Recall Customer Notification.
Initiated
2024-08-30
Root cause
Under Investigation by firm
Distribution
U.S. Nationwide distribution in the states of CA and VA.
Recall number
Z-0021-2025
Recall status
Open, Classified
CDRH record ID
209969
FDA classified
2024-10-07
Recall event number
95314

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗