Philips Medical Systems North America Co. Phillips
FDA classification date
Aug 31, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification
Reason stated in the source
1) During positioning of the Lateral Stand the rotation and/or skew movement may travel unexpectedly in opposite direction; 2) movement may be faster than normal speed; 3) the stand may not stop immediately upon release of the joystick.
Source facts
Action
A letter issued on 7/25/05 advising customers of problem. The letter advised that an upgrade will be performed.
Initiated
2005-07-27
Root cause
Other
Terminated
2008-05-20
Distribution
Devices were distributed to hospitals/medical centers nationwide.
Recall number
Z-0022-06
Recall status
Terminated
CDRH record ID
41206
FDA classified
2005-08-31
Recall event number
33014
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.