Philips had discovered through a field service testing where a device failed to comply with a Federal standard.
Source facts
Action
Philips Healthcare discovered through field service testing, that the device was in violation. Firm service rep. visited site on 7/01/14 and corrected unit.
The field service engineer immediately reprogrammed the generator for the X-ray tube installed in the system and re-adapted the x-ray tube to bring the system back into compliance.
At this time the Failure to comply has been corrected and the customer has been notified. This correction is to be free of charge to the customer.
For further questions please call (978) 687-1501.
Initiated
2014-08-08
Root cause
Radiation Control for Health and Safety Act
Terminated
2014-12-03
Distribution
US distribution to FL
Recall number
Z-0022-2015
Recall status
Terminated
CDRH record ID
129837
FDA classified
2014-10-09
Recall event number
69197