Z-0022-2018: Proclaim 5 Elite Implantable Pulse Generator and Proclaim 7 Elite Implantable Pulse Generator Models 3660

Recall record Z-0022-2018 for St. Jude Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0022-2018
Company in FDA record
St. Jude Medical, Inc.
FDA record creation date
Oct 17, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Analysis of the difference between the actual device longevity and the battery indicator status provided by the Clinician Programmer (CP) or Patient Controller (PC) has revealed errors in the CP or PC longevity estimations that can overestimate or under estimate the battery indicator status in certain situations.

Source facts

Action
St. Jude Medical sent an Important Medical Device Advisory letter dated September 12, 2017, to all affected customers. Customers were advised the following: Phase I: ERI Assessment If a patient reports that an ERI message is displayed on the PC, please contact your Abbott Representative to obtain the device-specific generator logs and to conduct the ERI Assessment using the steps in Appendix B prior to making an elective replacement decision. After the ERI Assessment is completed, follow-up communication from your Abbott representative will be provided in order to make appropriate decisions. The response will be one of two possible responses: o ERI is valid at this time for this device, follow existing information provided by IFU and CP for the elective replacement window for this device. o The device has the appropriate level of battery voltage to provide the existing therapy until the next assessment, which should occur no later than March 2018. During this period, the ERI message will be displayed on the PC at the start of each session. This message may be dismissed to enter the session on the PC. Phase II: ERI Software Update Abbott will deploy a software upgrade, upon approval by local governing agencies that addresses the errors in calculation causing the inappropriate ERI message. When this software update is deployed, the Phase I ERI assessment process is no longer needed. After the CP is updated, replacement decisions can be made following the revised instructions for use, which will be available at manuals.sjm.com. Once approved for use in your area, notification of the availability of this software update will occur through the Apple Public App Store for the PC and through the SJM App Catalog for the CP. Should you have questions about patient management, including questions regarding error messages related to IPGs, please contact your local Sales Representative or Abbott Support at 1-800-727-7846 (Opt3) (U.S) and +46 8 474 414
Initiated
2017-09-12
Root cause
Software design
Terminated
2020-04-16
Distribution
Worldwide Distribution - US (nationwide) (foreign consignee list will be provide by the firm at a later time)
Recall number
Z-0022-2018
Recall status
Terminated
CDRH record ID
158787
FDA record created
2017-10-17
Recall event number
78154

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗