Open original FDA record ↗Recall record
- FDA identifier
- Z-0023-2015
- Company in FDA record
- Biomedix, Inc.
- FDA classification date
- Oct 6, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Customer contacted Biomedix of complaint and returned the failed set along with 86 units of unopened stock (96 units originally shipped). Biomedix performed extensive investigation and found insufficient application of glue. Recall was initiated due to the rate of the failure.
Source facts
- Action
- Biomedix sent an PRODUCT RECALL NOTIFICATION letter dated August 14, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Please contact us with the quantity remaining in inventory or available to be returned to our facility. All units returned to Biomedix will be replaced with the same SELEC-3 item or a similar design at no charge. Please call (1-800-627-2765 ) for further questions.
- Initiated
- 2014-08-14
- Root cause
- Process control
- Terminated
- 2014-11-13
- Distribution
- US Distribution including the states of OH, NC, and TX
- Recall number
- Z-0023-2015
- Recall status
- Terminated
- CDRH record ID
- 130088
- FDA classified
- 2014-10-06
- Recall event number
- 69249