Skip to content

Public FDA record

Z-0023-2020: C2 Nerve Monitor, 4 Channel, 5140-508-240, UDI 07613327177329 Product Usage: The C2 Nerve Monitor Syste

Recall record Z-0023-2020 for Stryker Instruments Div. of Stryker Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0023-2020
Company in FDA record
Stryker Instruments Div. of Stryker Corporation
FDA record creation date
Oct 1, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Stryker C2 NerveMonitor's design and instructions may not optimally address usability issues related to the functionality of the device, which may result in use errors potentially causing or contributing to nerve injury.

Source facts

Action
On July 31, 2019, the firm distributed an Urgent Medical Device Recall Notification to its consignees to inform them of the recall. Customers were asked to do the following: 1. Immediately review the Recall Notification and this Business Reply Form (BRF). 2. Inform all users of the Stryker C2 NerveMonitor of the information contained in this recall notification. 3. Check all treatment areas, operating rooms and/or storage areas for affected products. 4. Determine a plan to transition to an alternative commercially available product. Document your transition date on the BRF. This transition date must be no later than January 31, 2020. 5. Complete page 2 of BRF to confirm receipt of this Notification, identify how many affected items are currently in your inventory, and provide the date you will transition to alternative product and return your Stryker C2 NerveMonitor. Upon receipt of your returned device(s), Stryker will work with your facility to issue a refund or credit. Please complete and return the BRF even if you dont have any affected product on hand. 6. Fax the completed BRF to Stryker Instruments at 8665212762, or email to kara.spath@stryker.com. 7. If you have further distributed this product, please forward copies of the Notification and this BRF to all affected locations, for each location to complete and return. Even if you have distributed all product to another location, please complete a BRF and indicate each location that received product.
Initiated
2019-07-31
Root cause
Device Design
Terminated
2021-03-24
Distribution
US Nationwide distribution.
Recall number
Z-0023-2020
Recall status
Terminated
CDRH record ID
175420
FDA record created
2019-10-01
Recall event number
83490

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗