Open original FDA record ↗Recall record
- FDA identifier
- Z-0024-2007
- Company in FDA record
- Abbott Diagnostic International, Ltd.
- FDA classification date
- Oct 12, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Following a customer complaint, Abbott found an atypical stability profile for the lot of the calibrators listed in this recall. The investigation to date has shown that both controls and patient results have shifted upwards over time together.
Source facts
- Action
- The Device Recall Letter dated September 07, 2006, and Customer Reply Form was sent via Federal Express Priority mail to all customers who received the affected lots.
- Initiated
- 2006-09-07
- Root cause
- Other
- Terminated
- 2011-09-09
- Distribution
- Worldwide, including USA, Puerto Rico, Canada, Mexico, Guatemala, Costa Rica, Venezuela, Colombia, Ecuador, Chile, Uruguay, Argentina, Brazil, Dominican Republic, El Salvador, Germany, Japan, Singapore, Hong Kong, Taiwan, Thailand, South Korea, Australia, Honduras, Trinidad-Tobago, St. Vincent, Bahamas, Panama, and Cayman Islands.
- Recall number
- Z-0024-2007
- Recall status
- Terminated
- CDRH record ID
- 48602
- FDA classified
- 2006-10-12
- Recall event number
- 36405