Open original FDA record ↗Recall record
- FDA identifier
- Z-0024-2015
- Company in FDA record
- Siemens Healthcare Diagnostics, Inc.
- FDA classification date
- Oct 6, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Siemens Healthcare Diagnostics has determined that Dimension Vista® Mass CKMB Isoenzyme Calibrator (MMB CAL) shows drift during routine stability monitoring that exceeds our acceptance criteria for drift over the shelf life of this product. Siemens has observed drift from 3% to 12% at MMB concentrations across the assay range. Depending on quality control limits, this drift may not have been d
Source facts
- Action
- An Urgent Medical Device Recall notice, dated August 2014, was provided to all Dimension Vista MMB CAL customers who received the affected lots to notify them of the issue, the potential risk to health, and actions to be taken by customers. Customers were instructed to discard lots 3JD045, 4BD064, 4CD012, and 4DD065, and recalibrate with lot 4FD073.
- Initiated
- 2014-08-28
- Root cause
- Process design
- Terminated
- 2015-03-30
- Distribution
- Worldwided distribution: US (nationwide) and countries of: Australia, Austria, Belgium, Canada, Denmark, Germany, Finland, France, Italy, Japan, Netherlands, New Zealand, Portugal, Saudi Arabia, Slovenia, South Korea, Spain, and Switzerland.
- Recall number
- Z-0024-2015
- Recall status
- Terminated
- CDRH record ID
- 129748
- FDA classified
- 2014-10-06
- Recall event number
- 69149